Directory of Tests:

MPV

Test Name

MPV

Test Code

1259

Unit

fL

Station

Hematology

Synonym Name/Abbreviation

Mean Platelet Value

Performing Facility

American Medical Labs, Herzliya
Pituah

Analyzer

Siemens Advia 2120 or 120

Test Method

Automated Cell Counter

AML Preferred Tube

3 mL EDTA whole blood

Storage Instructions

Room temperature up to 4h. Stored and transported
refrigerated up to 24 h.

Min. Volume

1 mL

 

Comments

Gently Invert tube 8-10 times after venipuncture to ensure
smooth mixing with anti-coagulant

Turn-Around-Time

One day

 

Intended Use

MPV is a sizing parameter, that may be abnormal in some
clinical situations (ITP). It can be used to make inferences about
platelet production in bone marrow.

Limitations

Clotted sample; hemolysis; tube not filled to correct volume.
Clumping may cause low platelets count.

Patient Preparation

No patient preparation
needed.

 

Reasons for Rejection

Clotted sample, improper tube, incorrect volume, incorrect
labeling, dilution of blood with I.V. fluid, samples older than 24
hours.

Adult Reference Ranges

Male and Female: 7.0 to 10.0
fL

Copyright © 2016. American Medical Laboratories. All rights reserved.